国产卡一卡二卡3卡四卡精品_mvp情人优酷_大学生无套带白浆嗯呢啊污_亚洲精品久久7777777国产_韩国伦理片在线2018_黄sei片网站_美女成人网_亚洲精品色情AAA片_亚洲网址在线_香蕉视频污在线观看_午夜精品久久久久久久久日韩欧美_四季在线观看免费资源_btv养生堂_亚洲国产中文在线视频_日本精品一区二区三区在线_野花高清在线观看免费高清_...中文天堂在线观看_亚洲精品久久久久久久不卡四虎_地心历险记1快播_暖暖 免费高清 日本

Welcome to the official website of Wuxi High Speed Analytical Instrument Co., Ltd.!
中文    English
Sales hotline+86 510 83202007
+86 510 83202173

High-frequency infrared carbon-sulfur meter pool voltage failures and solutions

Release time:2020-09-30 Views:1707 來源:本站

The cell voltage mainly has the following fault conditions:
1. Failure condition: The cell voltage of the carbon and sulfur analyzer fluctuates unsteadily, with fluctuations greater than 0.01V/min (normally it should be light rain 0.002V/min).
   Troubleshooting:
(1) If the signal of one channel of carbon and sulfur is unstable, it may be that the components of the amplification board of this channel are damaged, and the amplification board needs to be replaced.
(2) If both carbon and sulfur are unstable, it may be that the motor conversion is unstable. You can check whether the motor drive voltage is normal, and then check the motor drive and motor. If there is a problem, you may need to replace the motor drive board or motor.
2. The cause of the failure: the cell voltage signal of the carbon-sulfur analyzer is greater than 1.9V and remains unchanged.
   Troubleshooting: The cell voltage signal is too large and the display is over range. Adjust the 50K potentiometer of the amplified version and turn it down counterclockwise until the signal is stable at about 1.5000V.
3. The cause of the failure: the cell voltage has no signal, and the display is 0.0000V.
   Troubleshooting: No signal input, indicating that the signal from the lower computer cannot be transmitted to the upper computer, you can first check whether the USB is connected properly or the USB driver is installed correctly, and the USB cable needs to be adjusted.
4. The cause of the failure: the cell voltage signal is low, and the cell voltage is lower than 1.0000V.
   Troubleshooting: The cell voltage of the carbon-sulfur analyzer is small, but the beating is stable, because the inner wall of the gold-plated tube is dirty and the signal magnification is reduced. Need to contact the manufacturer to clean or replace the gold-plated tube.
5. The cause of the failure: the cell voltage signal is extremely low, less than 0.1000V.
   Troubleshooting:
(1) If the signal of one channel of carbon and sulfur is extremely low, the light source is not working. Check whether the light source has voltage and whether the light source is broken, and the light source needs to be replaced.
(2) Check whether the modulation motor is rotating. If the motor does not rotate, the cell voltage displays a zero signal, and the motor drive board or motor needs to be replaced.
主站蜘蛛池模板: 五台县| 甘德县| 北京市| 博客| 独山县| 新野县| 时尚| 玉林市| 黑龙江省| 贵港市| 山东| 靖江市| 武冈市| 玛曲县| 肇东市| 贵阳市| 新绛县| 红安县| 龙陵县| 闻喜县| 泗洪县| 沅陵县| 西和县| 延庆县| 黎川县| 常德市| 灵山县| 岗巴县| 高尔夫| 阳曲县| 辽中县| 龙江县| 开原市| 蒙阴县| 鹰潭市| 郯城县| 惠安县| 进贤县| 东海县| 永康市| 灌南县|